AccuMac AM176X (AM1760 / AM1762) Secondary SPRTs

Temperature range: −200 °C … 420 °C

Nominal resistance:100 ohms (model AM1760) or 25 ohms (model AM1762); coefficient α = 0.003925 ohms/ohm/°C

Accuracy (k=2):±0.010 °C at −196 °C; ±0.006 °C at 0.01 °C; ±0.015 °C at 420 °C; ±0.025 °C at 661 °C

Stability and drift:Short-term instability <0.002 °C; drift in the TTV <0.004 °C (4 μK) after 100 hours at 661 °C

Thermal shock and self-heating: Thermal shock <0.002 °C (10 min-max cycles); self-heating 0.0015 °C (at 0.5 mA for 100 Ω / 1 mA for 25 Ω); response time 9 s

Material of the protective sleeve:Inconel™ heat-resistant alloy; diameter 6.35 mm (0.25″); standard length 305 mm (12″) or 508 mm (20″); BEND option

Connection and calibration:A 2-metre-long 4-core Teflon cable; metal handle (up to 180 °C); Lemo Smart Connector, spade and banana plugs

TEKKNOW

Request a callback

Contact us for a free consultation and a quote for our services.

    The AM1760 series Secondary SPRTs provides our customers with reliable secondary standards that can be used daily in their labs. These SPRTs feature an accuracy of <0.006°C at 0.01°C, a short term stability of <0.002°C and a very low drift rate of less than 0.004°C after 100 hours at 661°C. Two different lengths of these SPRTs are available at either 12 inches (305 mm) or 20 inches (508 mm).

    The sensing element is designed to protect the platinum sensing wire from contamination at high temperatures, giving these SPRTs a high level of stability and repeatability in performance. A uniquely designed support structure and filling material provides excellent balance between the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard with an ISO 17025 accredited calibration. Available in 100 Ω (AM1760) and 25 Ω (AM1762) nominal resistances.

     

    Model

    AM176X

    Range

    −200 °C … 670 °C

    Accuracy

    ±0.006 °C at 0.01 °C

    ,

    ±0.010 °C at −196 °C

    ,

    ±0.015 °C at 420 °C

    ,

    ±0.025 °C at 661 °C

    Stability

    drift in TTV <0.004 °C (4 μK) after 100 hours at 661 °C

    ,

    Short-term instability <0.002 °C

    Heatstroke

    <0.002 °C (10 min-max cycles)

    ,

    self-heating: 0.0015 °C (at 0.5 mA for 100 Ω / 1 mA for 25 Ω); response time: 9 s

    Dimensions

    diameter 6.35 mm (0.25"); standard length 305 mm (12") or 508 mm (20")

    Connection

    A 2-metre-long 4-core Teflon cable; metal handle (up to 180 °C); Lemo Smart Connector, spade and banana plugs

    We offer automatic pressure calibrators, digital pressure gauges, pressure calibration pumps, and multifunctional process calibrators

    DataSheet AM1760_and_AM1762_Secondary_Sprts