Intrinsically Safe Dew-Point Transmitter — Michell Easidew I.S.

Measurement Range: -100 to +20 °Cdp (-148 to +68 °Fdp) or -110 to +20 °Cdp (-166 to +68 °Fdp); moisture content 0 to 3000 ppmV

Accuracy & Repeatability: ±2 °Cdp (±3.6 °Fdp) over compensated temperature range (-20 to +50 °C); repeatability 0.5 °Cdp

Response time: 5 mins to T95 (dry to wet)

Calibration: Traceable 7-point calibration certificate (NPL / NIST)

Output Signal: 4 to 20 mA (2-wire loop-powered, current source, user configurable); diagnostics: fault 23 mA, under-range 4 mA, over-range 20 mA

Supply Voltage & Load: 12 to 28 V DC; max load 250 Ω @ 12 V (500 Ω @ 24 V); current consumption 20 mA max (23 mA fault)

Hazardous Area Approvals: ATEX/UKCA: II 1 G Ex ia IIC T4 Ga (-20 to +70 °C)
IECEx: Ex ia IIC T4 Ga (−20 … +70 °C)
TR CU: 0Ex ia IIC T4 Ga (-20 to +70 °C)
cQPSus: IS, Class I, Div 1, Groups A, B, C & D, T4; Class I, Zone 0, AEx ia IIC T4 Ga
Approved isolators: KFD0-CS-EX1.50P, KFD0-CS-EX2.50P, KFD2-STC4-EX1.H

Operating Pressure & Flow: 52.5 MPa (525 barg / 7614 psig) max; overpressure rating 90 MPa (900 barg); flow 1 to 5 Nl/min (block), 0 to 10 m/s (direct insertion)

TEKKNOW

Request a callback

Contact us for a free consultation and a quote for our services.

    The Michell Easidew I.S. is an intrinsically safe 2-wire dew-point transmitter certified for hazardous area applications in flammable or explosive gases and non-polar liquids, carrying ATEX, IECEx, UKCA, cQPSus, and TR CU Ex (Zone 0) approvals.

    Built around Michell's Ceramic Metal-Oxide Moisture Sensor in a 316 stainless steel IP66 body rated to 525 barg (900 barg overpressure), it measures down to -110 °Cdp with ±2 °Cdp accuracy and is delivered with a traceable calibration certificate.

    Choice of 5/8" UNF, G1/2" BSP, or 3/4" UNF process connections enables straightforward installation or retrofit, and it pairs directly with the Michell Intrinsically Safe Field Display and Easidew & Pura Communications Kit.

    Model

    Michell Easidew I.S.

    Range

    -100 to +20 °Cdp (-148 to +68 °Fdp) or -110 to +20 °Cdp (-166 to +68 °Fdp); moisture content 0 to 3000 ppmV

    Accuracy

    ±2 °Cdp (±3.6 °Fdp) over compensated temperature range (-20 to +50 °C); repeatability 0.5 °Cdp

    Communication

    4 to 20 mA (2-wire loop-powered, current source, user configurable); diagnostics: fault 23 mA, under-range 4 mA, over-range 20 mA

    Power supply

    12 to 28 V DC; max load 250 Ω @ 12 V (500 Ω @ 24 V); current consumption 20 mA max (23 mA fault)

    Protection/Application area

    ATEX/UKCA: II 1 G Ex ia IIC T4 Ga (-20…+70 °C) IECEx: Ex ia IIC T4 Ga (-20…+70 °C) TR CU: 0Ex ia IIC T4 Ga (-20…+70 °C) cQPSus: IS, Class I

    Connection

    5/8"-18 UNF (alternatives: G1/2" BSP or 3/4"-16 UNF), 316 stainless steel

    Features

    5 mins to T95 (dry to wet)

    We develop and manufacture reference equipment for laboratories and metrology, as well as precision dew-point analyzers, hygrometers, and humidity sensors for industrial processes, gas drying, and environmental monitoring.

    Technical Documentation: Michell_Instruments_Easidew_TX_I.S._Datasheet