Transmille 3010A Precision Multi-Product Calibrator
Direct current voltage (DCV): 0 … ±1000 V; accuracy of ±100 ppm (0.01%)
Alternating current voltage (ACV): 1 mV … 1,000 V (10 Hz … 100 kHz); accuracy: ±0.05%
Direct Current Intensity (DCI): 0 … ±30 A (up to 1,500 A with the EA002 current coil); accuracy: ±0.02%
Alternating current (ACI): 10 µA … 30 A (10 Hz … 10 kHz) (up to 1,500 A with a coil); accuracy: ±0.07%
Resistance: 0 Ω … 1 GΩ (from ±50 ppm)
Capacity: 1 nF … 10 mF
Frequency: 1 Hz … 10 MHz
Power (PWRDDS option): AC/DC power calibration up to 30 kW with independent adjustment of current, voltage and phase shift
Oscilloscope (optional): Calibration of oscilloscopes for amplitude, time marks and bandwidth up to 250 MHz / 1 GHz
Ergonomics and controls: Two informative LCD displays, an ergonomic encoder, and digital interfaces including USB, RS-232 and GPIB
Request a callback
Contact us for a free consultation and a quote for our services.
The Transmille 3010A is an entry-level precision multi-product calibrator offering an economical yet comprehensive calibration solution with 100 ppm base DC accuracy. Designed to cover the widest workload of instruments in industrial calibration laboratories, the 3010A calibrates up to 18 distinct instrument types, bridging the gap between older analog meters and modern 3.5 to 4.5 digit digital multimeters.
Equipped with 30A current drive as standard (extendable to 1500A with current coils), 1000V AC/DC output, 1 GΩ resistance, and 10 mF capacitance, the 3010A can be expanded with internal and external options including oscilloscope calibration, AC/DC power, temperature simulation, and insulation test calibration. Fully automated with Transmille ProCal software.
| Model |
3010A |
|---|---|
| DCV voltage |
0 … ±1,000 V ,accuracy ±100 ppm (0.01 per cent) |
| ACV voltage |
1 mV … 1,000 V (10 Hz … 100 kHz) ,accuracy ±0.05% |
| DCI current |
0 … ±30 A, accuracy of ±0.02% ,up to 1,500 A with the EA002 current coil |
| ACI current |
10 µA … 30 A (10 Hz … 10 kHz), accuracy: ±0.07% ,up to 1,500 A with the EA002 coil |
| Resistance |
0 Ω … 1 GΩ (from ±50 ppm) |
| Capacity: |
Capacitance: 1 nF … 10 mF |
| Frequency |
1 Hz … 10 MHz |
| Power |
AC/DC up to 30 kW with independent control of current, voltage and phase shift |
| Oscilloscope |
Calibration of oscilloscopes for amplitude, time marks and bandwidth up to 250 MHz / 1 GHz (optional) |
DataSheet 3010A_Extended_Specifications