Transmille 3041A Advanced Precision Multi-Product Calibrator

Direct current voltage (DCV): 0 … ±1000 V; basic accuracy over 1 year of ±25 ppm (0.0025%)

Alternating current voltage (ACV): 1 mV … 1,000 V (10 Hz … 500 kHz); accuracy: ±0.035%

Direct Current Intensity (DCI): 0 … ±30 A (up to 1,500 A with the EA002 coil); accuracy of ±50 ppm (0.005%)

Alternating current (ACI): 10 µA … 30 A (10 Hz … 10 kHz) (up to 1,500 A with a coil); accuracy: ±0.04%

Resistance: 0 Ω … 1 GΩ (passive precision decade boxes and continuous adjustment); accuracy of ±10 ppm

Capacity: 1 nF … 10 mF (±0.2%)

Inductance: 1 mG … 10 G (IND option)

Power (PWRDDS option): AC/DC power up to 30 kW with a phase shift of 0 … 360°, simulation of harmonics and power quality

Oscilloscope (optional): 250 MHz, 1 GHz and 3 GHz options (fast transient response <100 ps, time markers, amplitude up to 50 V/div)

TEKKNOW

Request a callback

Contact us for a free consultation and a quote for our services.

    The Transmille 3041A is the flagship multi-product calibrator in the 3000A Series, engineered for high-precision calibration laboratories demanding superior accuracy (25 ppm) and maximum workload capability. The 3041A provides exceptional coverage for multimeters up to 6.5 and 7.5 digits, wideband AC instruments, power analyzers, clamp meters, and high-frequency oscilloscopes.

    With standard 30A current output, 1000V AC/DC, 1 GΩ resistance, and 10 mF capacitance, the 3041A accommodates an unparalleled array of internal options (including 3 GHz oscilloscope calibration, AC/DC power/harmonics, inductance, and RTD/thermocouple simulation). Its dual-display design and full compatibility with ProCal automated calibration software make it the premier choice for national, defense, and commercial calibration centers worldwide.

    Model

    3041A

    DCV voltage

    0 … ±1,000 V

    ,

    0 … ±1000 V; basic accuracy over 1 year of ±25 ppm (0.0025%)

    ACV voltage

    1 mV … 1,000 V (10 Hz … 500 kHz)

    ,

    accuracy ±0.035%

    DCI current

    0 … ±30 A, accuracy of ±50 ppm (0.005%)

    ,

    up to 1,500 A with the EA002 current coil

    ACI current

    10 µA … 30 A (10 Hz … 10 kHz), accuracy: ±0.04%

    ,

    up to 1,500 A with the EA002 coil

    Resistance

    0 Ω … 1 GΩ (passive precision decade boxes and continuous adjustment); accuracy of ±10 ppm

    Capacity:

    1 nF … 10 mF (±0.2%)

    Power

    AC/DC power up to 30 kW with a phase shift of 0 … 360°, simulation of harmonics and power quality

    Oscilloscope

    250 MHz, 1 GHz and 3 GHz options (fast transient response <100 ps, time markers, amplitude up to 50 V/div)

    We develop and manufacture high-precision calibrators, reference standards, and specialized software for automating measurements, metrological control, and instrument verification in laboratories.