AccuMac AM1710 Secondary Reference PRT

Temperature range: −200 °C … 160 °C

Nominal resistance: 100 ohms at 0.01 °C (temperature coefficient α = 0.003925 ohms/ohm/°C, pure platinum in accordance with MTSh-90)

Calibration error: ±0.025 °C at −40 °C; ±0.015 °C at 0.01 °C; ±0.025 °C at 160 °C (accuracy <0.012 °C at 0.01 °C)

Stability and drift: Short-term stability <0.007 °C; long-term drift <0.01 °C in the temperature-time curve after 100 hours at 160 °C

Thermal shock and hysteresis: Thermal shock <0.005 °C (10 thermal cycles); hysteresis ≤0.005 °C; self-heating: 50 mW/°C; response time: 9 s

Response time: 4 seconds (for 63 per cent of the transient process in water at a flow velocity of 1 m/s)

Design and dimensions of the sleeve: Stainless steel; straight sleeve AM1710-12 (length 305 mm / 12″, diameter 6.35 mm / 1/4″) or 90° bent sleeve AM1710-BEND

Cable and connection: 2-metre 4-core Teflon cable (Lemo Smart Connector, spade terminals or banana plugs)

Calibration: Supplied with an ISO 17025-accredited calibration certificate

TEKKNOW

Request a callback

Contact us for a free consultation and a quote for our services.

    The AM1710 series of Secondary Reference PRTs offers affordable reference probes for precision temperature measurement and calibration both in the lab and in the field. These PRTs feature an accuracy of <0.012°C and a short term stability of <0.007°C across -40°C to 160°C.

    To reach the best performance in stability and repeatability, the sensing element has been specially designed to protect the platinum sensing wire from contamination at high temperatures. A unique supporting structure and filling material provide the best balance among the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard with an ISO 17025 accredited calibration. Available in straight (AM1710-12) and bent (AM1710-BEND) configurations.

    Model

    AM1710

    Range

    −40 °C … 160 °C

    Stability

    Long-term drift of <0.01 °C in the TTV after 100 hours at 160 °C

    ,

    Short-term stability <0.007 °C

    Hysteresis

    ≤0,005 °C

    Heatstroke

    <0.005 °C (10 thermal cycles)

    Dimensions

    AM1710-12 straight sleeve (length 305 mm / 12", diameter 6.35 mm / 1/4") or AM1710-BEND 90° bend

    Connection

    A 2-metre-long 4-core Teflon cable; metal handle (up to 180 °C); Lemo Smart Connector, spade and banana plugs

    We offer automatic pressure calibrators, digital pressure gauges, pressure calibration pumps, and multifunctional process calibrators

    DataSheet AM1710_Secondary_Reference_Prts