Additel ADT286 Multifunction Reference Thermometer Readout
Error in the resistance ratio (Rx/Rs):1 ppm (channel 1 with external reference Rs); resistance accuracy up to 15 ppm (0.35 mΩ at 100 Ω)
Measurement channels: 2 basic precision channels (CH1/CH2); expandable to 82 channels using scanning modules (up to 10 ch/s)
Supported sensors: SPRTs, RTDs (PRTs: Pt100, Pt25, Pt10, Cu, Ni), thermistors, 15 types of thermocouples (S, R, B, K, N, E, J, T, etc.)
Built-in multimeter: 8½-digit DC multimeter (DC voltage from 100 nV, current, resistance)
Cold Joint Compensation (CJC): Accuracy: ±0.1 °C (1 year, 23 ± 5 °C); 10 CJC precision sensors on each 20-channel module
Automatic compensation for self-heating: Auto zero-power self-heating compensation
Interfaces and display: 10.1″ colour touchscreen LCD (1280×800); Wi-Fi, Ethernet (RJ-45), USB-A, USB-B; 10 GB built-in storage
Dimensions and weight:250 × 420 × 200 mm; weight: 8.39 kg (18.5 lbs); power supply: 100–240 V AC
Request a callback
Contact us for a free consultation and a quote for our services.
2019 IDEA Award Winner.
The Additel 286 Multifunction Reference Thermometer Readout is an industry first! We have combined the capabilities of a high-end reference thermometer with a highly capable data acquisition system and 8.5 digit multimeter. The ADT286 is capable of scanning and recording up to 82 channels at 10 channels per second. Users can easily configure the ADT286 to perform field calibrations and uniformity studies as well as use the unit in the lab as a precision thermometer and 8.5 digit multimeter. Get more for less with this newest game changer from Additel!
Scanner Modules & Expandability:
The base unit comes with two precision readout channels that can be used to measure your SPRT. Need to calibrate RTDs, PRTs, thermistors or thermocouples? Add a scanner module and you now have the ability to measure 10 RTDs, PRTs, or thermistors and 20 thermocouples. Expand up to 82 channels with our unique easy to use scanner modules. Each 20 channel module is outfitted with our proprietary universal terminals with an industry's best cold junction capability second to none. Utilize the module docked atop the ADT286, or connect remotely with cables to suit nearly any unique setup/configuration. Additel also has a process module specifically designed to measure process instrumentation like transmitters and switches, supplying loop power for transmitters.
Sensor Library & Automation:
Designed to make your job easier, the ADT286 has a large sensor library supporting 15 TC types, both standard and special limits, 18 different thermal resistors, CVD, ITS-90, and a large variety of standard curves for RTDs and thermistors. Loaded with special applications such as probe calibration, SPRT calibration, chamber mapping and more. Preinstalled drivers control Additel and other manufacturers' heat sources automatically.
| Model |
ADT286 |
|---|---|
| Accuracy |
1 ppm (channel 1 with external reference Rs) ,up to 15 ppm (0.35 mΩ at 100 Ω) |
| Датчики |
SPRTs, RTDs (PRTs: Pt100, Pt25, Pt10, Cu, Ni), thermistors, 15 types of thermocouples (S, R, B, K, N, E, J, T, etc.) |
| Features |
2 basic precision channels (CH1/CH2); expandable to 82 channels using scanning modules (up to 10 ch/s) ,8½-digit DC multimeter (DC voltage from 100 nV, current, resistance) ,Zero-power measurement function |
| Communication |
Ethernet ,USB-A ,USB-B ,WiFi |
| Dimensions |
250 × 420 × 200 mm |
| Weight |
8.39 kg |