Additel ADT850-1200 Laboratory Thermocouple Calibration Furnace
Temperature range: 100 °C … 1210 °C
Stability: ±0.1 °C full range of 100 °C … 1210 °C
Uniformity: Radial uniformity: ±0.25 °C at 1200 °C
Axial uniformity: ±0.25 °C over a 60 mm section at 1200 °C
3-in-1 configuration: 9 unique operating modes (optimised settings for short sensors, long thermocouples and annealing)
Dimensions and weight: 680 × 330 × 450 mm, weight ~35 kg
Control and heating: Multi-zone independent heating with an adaptive microprocessor-based control algorithm
Rapid cooling: ‘Quick Cool Technology’ active cooling system to reduce the transition cycle between points
Sensor positioning: Integrated sliding probe holder for precise depth adjustment and secure mechanical retention
Request a callback
Contact us for a free consultation and a quote for our services.
Thermocouple calibration work can be challenging. Traditional furnace designs require several individual devices to meet industry standards for various calibration applications. To address this costly reality, Additel has created a multi-purpose horizontal furnace to help save time, money and space in your calibration facility. Our new ADT850 Laboratory Thermocouple Calibration Furnace is like having three separate furnaces in one.
Users can select optimized settings for shorter probes, longer probes, and even annealing purposes. The ADT850 horizontal furnace can be used in (9) different modes/configurations to help meet even the most challenging calibration requirements and standards. Additel’s 850 furnace is packed with advanced features and a performance level you will not find anywhere else. The ADT850 is commonly used in a multitude of industries such as energy, calibration laboratories, aerospace, and metallurgy. It is generally used by primary and secondary calibration laboratories to calibrate various length noble and base metal thermocouples with the lowest possible uncertainties.
Key Highlights:
• 3-in-1 Design: 9 distinct modes for short probes, long probes, and annealing.
• Temperature Range: 100°C to 1200°C with stability of ±0.1°C and axial/radial uniformity of ±0.25°C at 1200°C.
• Quick Cool Technology: Drastically cuts turnaround times between test points.
• Sliding Probe Holder: Provides mechanical stability and precise, repeatable depth control for multiple probes simultaneously.
• Patent-Pending EMF Shielding: Eliminates stray electric fields and induced EMF noise for clean thermoelectric measurements.
• Pivoting Color Touchscreen: Intuitive modern interface with remote control via Wi-Fi/LAN.
| Model |
ADT850-1200 |
|---|---|
| Range |
100 °C … 1210 °C |
| Uniformity |
Axial uniformity: ±0.25 °C over a 60 mm section at 1200 °C ,Radial: ±0.25 °C at 1200 °C |
| Dimensions |
680 × 330 × 450 mm |
| Weight |
~35 kg |
DataSheet 850_Thermocouple_Calibration_Furnaces