Additel ADT87X-1210 Thermocouple Calibration Furnaces

Temperature range: 100 °C … 1210 °C

Accuracy: ADT878-1210 (Reference): display accuracy ±1.5 °C across the entire range
ADT875-1210 (Standard): display accuracy of ±2.0 °C across the entire range

Stability: ±0.1 °C full range of 100 °C … 1210 °C

Uniformity: Axial (over a 60 mm section from the bottom of the insert): ±0.25 °C
Radial: ±0.25 °C at 1200 °C

Bath dimensions and capacity: immersion depth up to 220 mm

Dimensions and weight: 395 × 205 × 400 mm; weight: approximately 16.5 kg

Heating zones and control: 3-zone patented adaptive control (‘Advanced Adaptive Control’) with wind tunnel control technology

Process calibrator (PC option):

4 built-in measurement channels:
• Reference thermocouple channel
• 3 channels for testing sensors (mA, V, switch test, thermocouples, HART, 24 V DC)
• Quick-Push quick-connect terminals for all types of thermocouples

TEKKNOW

Request a callback

Contact us for a free consultation and a quote for our services.

    We understand the many challenges associated with thermocouple calibration work. That is precisely why we decided to introduce the ADT875-1210 and ADT878-1210 Thermocouple Calibration Furnaces.

    With unmatched stability, uniformity, and an optional on-board process calibrator, calibrating thermocouples has never been easier. With two separate units to choose from, the ADT875-1210 (Standard) and ADT878-1210 (Reference) furnaces include a patented multi-zone temperature control which provides a never before seen, highly stable and uniform heat source to ensure you get the best possible results from a modest investment. With metallic interchangeable inserts, users have the flexibility needed to service a wide variety of units under test (UUT) and the durability they have come to expect from Additel. The furnaces can be purchased with or without our on-board process calibration electronics.

    Key Highlights:
    • Advanced Adaptive Control: Incorporates patent-pending wind tunnel control technology with Additel’s impressive 3-zone temperature control to achieve optimal axial and radial uniformity.
    • Temperature Range: 100°C to 1210°C with display accuracy of ±1.5°C (ADT878) or ±2.0°C (ADT875).
    • Process Calibrator Option (PC): Includes patented Quick-Push connectors accommodating virtually all TC connection types. Offers 4 measurement channels: reference thermocouple channel plus up to 3 test channels with mA, V, switch test, 24V loop power, HART protocol communication, and task documentation.
    • External Sensor Control: Allows controlling furnace temperature directly from an external control thermocouple, further reducing measurement uncertainties.

    Model

    ADT87X-1210

    Range

    100 °C … 1210 °C

    Accuracy

    ±1.5 °C full range

    ,

    ±2.0 °C full range

    Uniformity

    Axial (over a 60 mm section from the bottom of the insert): ±0.25 °C

    ,

    Radial: ±0.25 °C at 1200 °C

    Size and volume

    immersion depth up to 220 mm

    Dimensions

    395 × 205 × 400 мм

    Weight

    16.5 kg

    We offer automatic pressure calibrators, digital pressure gauges, pressure calibration pumps, and multifunctional process calibrators

    DataSheet 875-1210_and_878-1210_Thermocouple_Calibration_Furnaces