Process Moisture Analyzer — Michell Promet EExd
Channels & Sensor: Single- or dual-channel industrial analyzer; Michell Ceramic Metal-Oxide Moisture Sensor; sour gas & up to 20 % H2 compatible
Measurement & Pressure Range: -120 to +30 °Cdp (calibrated -100 to +20 °Cdp), 1 ppbV to 30,000 ppmV; operating pressure 30 to 138 barg (sensor to 250 barg)
Accuracy & Conversions: ±1 °Cdp (-59.9 to +20 °Cdp) / ±2 °Cdp (-100 to -60 °Cdp), ±10/20 % moisture content; pressure ±0.25 % FS; ISO 18453 & IGT#8 (NPL/NIST)
Connections & Flow: 1/8″ NPT(F) sample connections; sample flow 1 Nl/min (2.1 scfh); optional sample flow alarm
Outputs, HMI & Data Logging: Vacuum fluorescent display w/ capacitive through-glass touch screen; 2× 4–20 mA/ch (500 Ω), RS485 Modbus RTU, 2 relays/ch; 150 rolling logs
Power & Ambient Range: 90 to 260 V AC, 50/60 Hz, 180 W; internal temperature control; -20 to +60 °C (-4 to +140 °F), max 95 % RH (shaded)
Hazardous Area Approvals: ATEX/IECEx/UKCA II 2 G Ex db IIB+H2 Gb T5/T4/T3; cQPSus Cl I Div 1 Gr BCD / Zone 1; TR CU 012 (EAC), GOST-R, GOST-K
Enclosure Construction: Explosionproof/Flameproof EExd cast copper-free aluminium, polyester powder coated (BS3900); integrated hazardous-area sampling system
Request a callback
Contact us for a free consultation and a quote for our services.
The Michell Promet EExd Process Moisture Analyzer is a hazardous-area rated, single- or dual-channel industrial hygrometer system engineered for direct installation in Zone 1 / Class I Div 1 areas across offshore gas platforms, transmission pipelines, glycol dehydration units, and petrochemical facilities (compatible with sour gas and up to 20 % hydrogen).
Combining Michell ceramic metal-oxide sensors (calibrated traceable to NPL and NIST over -120 to +30 °Cdp / 1 ppbV to 30,000 ppmV) with integrated sample conditioning and pressure transducers (±0.25 % FS), it operates directly at 30 to 138 barg (sensor rated to 250 barg) and calculates moisture content per ISO 18453 or IGT#8.
Inside an internally temperature-controlled, polyester-coated aluminium EExd enclosure (-20 to +60 °C ambient), the analyzer offers a through-glass VFD touchscreen, two 4–20 mA outputs and two relays per channel, RS485 Modbus RTU, and global ATEX, IECEx, UKCA, cQPSus, TR CU 012, and GOST approvals.
| Model |
Michell Promet EExd |
|---|---|
| Range |
-120 to +30 °Cdp (1 ppbV to 30,000 ppmV); pressure 30 to 138 barg |
| Accuracy |
±1 °Cdp (-59.9 to +20 °Cdp) / ±2 °Cdp (-100 to -60 °Cdp); pressure ±0.25 % FS |
| Communication |
Through-glass VFD, 2× 4–20 mA/ch, RS485 Modbus RTU, 2 relays/ch |
| Power supply |
90 to 260 V AC, 50/60 Hz, 180 W |
| Protection/Application area |
ATEX/IECEx/UKCA Ex db IIB+H2 Gb T5/T4/T3, cQPSus Cl I Div 1, TR CU 012 |
| Connection |
Gas 1/8″ NPT(F), flow 1 Nl/min, integrated sampling system |
| Features |
1–2 channels (dew point & pressure), sour gas & 20 % H2 ready, Zone 1 mount |
Technical Documentation: Michell_Instruments_Promet_EExd_EN_Datasheet