Ultra-Pure Moisture Hygrometers — Michell Pura Online / Pura Advanced Online 2

Measurement Range: -120 to -40 °Cdp (-184 to -40 °Fdp), down to 1 ppbV; 0 to 127 ppmV (ppmV, ppbV, lbs/MMscf, g/m³ on Advanced Online 2)

Accuracy: ±1 °C (-40 to -60 °Cdp); ±2 °C (-60 to -100 °Cdp); ±4 °C (-100 to -120 °Cdp, extrapolated)

Response time: Fast response to moisture breakthrough (ceramic metal-oxide sensor; exact T95 not stated in datasheet)

Calibration: Traceable 7-point calibration certificate (NPL / NIST; UKAS ISO/IEC 17025 on request)

Output Signals & Relays: Pura Advanced Online 2 (5.5-digit touchscreen + live pressure compensation): 3 × 4–20 mA, Modbus RTU RS485, 4 NAMUR relays (30 V DC 5 A)
Pura Online (4-digit economical monitor): 1 × 4–20/0–20 mA, Modbus RTU RS485, 2 relays (3 A / 5 A @ 250 V AC)

Supply Voltage: Advanced Online 2: 85–265 V AC (7.5 VA) or 18–28 V DC (170 mA @ 24 V DC)
Pura Online: 85–265 V AC or 24 V DC (60 mA max)

Hazardous Area & Approvals: Optional Pura I.S. sensor: ATEX/UKCA II 1 G Ex ia IIC T4 Ga, IECEx Ex ia IIC T4 Ga, TR CU 0Ex ia IIC T4 Ga (-20…+70 °C), cQPSus Class I Div 1 Gr. A–D T4 / Zone 0 AEx ia IIC T4 Ga; Monitor: BS/EN 61010-1:2010

Operating Pressure & Flow: 10⁻⁷ Pa (10⁻⁹ torr) min to 24 MPa (240 barg / 3481 psig) max; flow 1 to 5 Nl/min (block) or 0 to 10 m/s (direct)

TEKKNOW

Request a callback

Contact us for a free consultation and a quote for our services.

    The Michell Pura Online and Pura Advanced Online 2 hygrometers deliver economical, stable trace moisture measurement down to -120 °Cdp (<1 ppbV) for gas purification, semiconductor fabrication, fiber optics, optical coating, and specialty gas distribution.

    Available with either the touchscreen Pura Advanced Online 2 monitor (live pressure compensation, 3 × 4–20 mA, Modbus RTU, 4 NAMUR relays) or the economical Pura Online monitor (1 × 4–20/0–20 mA, Modbus RTU, 2 relays), the system pairs with an electropolished 0.1–0.2 Ra µm 316 SS VCR sensor.

    Optional Pura I.S. hazardous area certification, UHP double-bagged cleanroom packaging, and Michell's Sensor Exchange program ensure maximum reliability and minimal maintenance cost.

    Model

    Michell Pura Online / Advanced Online 2

    Range

    -120 to -40 °Cdp (-184 to -40 °Fdp), down to 1 ppbV; 0 to 127 ppmV (ppmV, ppbV, lbs/MMscf, g/m³ on Advanced Online 2)

    Accuracy

    ±1 °C (-40 to -60 °Cdp); ±2 °C (-60 to -100 °Cdp); ±4 °C (-100 to -120 °Cdp, extrapolated)

    Communication

    Pura Advanced Online 2 (5.5-digit touchscreen + live pressure compensation): 3 × 4–20 mA, Modbus RTU RS485

    Power supply

    Advanced Online 2: 85–265 V AC (7.5 VA) or 18–28 V DC (170 mA @ 24 V DC) Pura Online: 85–265 V AC or 24 V DC (60 mA max)

    Protection/Application area

    Optional Pura I.S. sensor: ATEX/UKCA II 1 G Ex ia IIC T4 Ga, IECEx Ex ia IIC T4 Ga, TR CU 0Ex ia IIC T4 Ga (-20…+70 °C)

    Connection

    2 × 1/4" male VCR (PRM/OEM, electropolished 0.1–0.2 Ra µm) or 1/2" male VCR (SEN)