Process Moisture Analyzer — Michell QMA601
Technology & Range: Fast-Response Quartz Crystal Microbalance (QCM); calibrated 0.1 to 700 ppmV (measuring 0.1 to 2000 ppmV; LoD 0.1 ppmV); up to 20 % H2 ready
Accuracy & Repeatability: ±0.1 ppmV (0.1–1 ppmV) / ±10 % (1–2000 ppmV); repeatability ±0.1 ppmV / ±5 %; internal NPL/NIST traceable moisture generator
Response & Carrier Gases: T63 <2 min, T95 <5 min in either direction; sensitivity 0.01 ppmV or 1 %; 21 predefined + 3 user-defined carrier gases
Pressure & Sample Flow: Inlet pressure 2 barg, outlet 1 barg; total sample flow 300 ml/min; sample gas temp 0 to +100 °C
HMI, Outputs & Alarms: 7″ through-the-glass color touchscreen LCD (no hot-work permit needed), 2× 4–20 mA / 1–5 V, RS485 Modbus RTU, Modbus TCP, 4 Form C relays, 14 alarms
Power & Ambient Range: 85 to 264 V AC (47/63 Hz) or 24 V DC (ATEX/IECEx; 24 V DC only cQPSus); +5 to +45 °C (-20 to +55 °C w/ sampling cabinet)
Hazardous Area Approvals: ATEX/UKCA/IECEx II 2 GD Ex db IIB+H2 T6 Gb (-40 to +60 °C); cQPSus Cl I Div 1 Gr BCD T6; TR CU 1Ex d IIB+H2 T4/T3 Gb X; KZ approval
Package, Connections & Weight: Flameproof GUB cast copper-free aluminium LM25 IP66 / NEMA 4; 316L sampling system; 1/8″ NPT(F), M20; weight 35 kg
Request a callback
Contact us for a free consultation and a quote for our services.
The Michell QMA601 Process Moisture Analyzer delivers fast, reliable trace moisture measurements from 0.1 to 2000 ppmV in variable carrier gases—including natural gas dehydration, up to 20 % hydrogen blends, ethylene/propylene production, and catalytic reforming recycle streams—using next-generation Quartz Crystal Microbalance (QCM) sensing.
An internal moisture generator traceable to NPL and NIST performs built-in verification on customer process gas, allowing up to 3 years of maintenance-free operation. Fast reversible sorption cycling achieves T63 <2 minutes and T95 <5 minutes in both wet-to-dry and dry-to-wet directions with ±0.1 ppmV sub-ppm accuracy.
Certified globally (ATEX/UKCA/IECEx Ex db IIB+H2 T6 Gb, cQPSus Cl I Div 1, TR CU 012, and Kazakhstan pattern approval) in a cast aluminium IP66 / NEMA 4 Ex d housing (35 kg), the QMA601 provides a 7″ color HMI operated through the glass without a hot-work permit, 2× 4–20 mA / 1–5 V outputs, RS485 Modbus RTU, Modbus TCP, and 21 predefined gas profiles.
| Model |
Michell QMA601 |
|---|---|
| Range |
0.1 to 2000 ppmV (calibrated 0.1 to 700 ppmV); pressure 2 barg |
| Accuracy |
±0.1 ppmV (<1 ppmV) / ±10 % of reading; NPL/NIST self-cal |
| Communication |
7″ through-glass LCD, 2× 4–20 mA / 1–5 V, Modbus RTU/TCP, 4 relays |
| Power supply |
85 to 264 V AC (47/63 Hz) or 24 V DC |
| Protection/Application area |
IP66 / NEMA 4 Ex d; ATEX/IECEx Ex db IIB+H2 T6 Gb, cQPSus, TR CU |
| Connection |
Gas 1/8″ NPT(F), M20 cable glands, 316L sampling system |
| Features |
QCM sensor, T95 <5 min response, 3-yr maintenance free, 35 kg |
Technical Documentation: Michell_Instruments_QMA601_Datasheet